Hi guys
I have a Fluke 789 I have recently brought second hand that simulates 0-20mA instead of 4-20. Is there a fault in the meter or can you change it back to 4-20??. Have not seen it before.
Also note that the source function does not work.
Cheers
Dave
Bit of a late reply sorry.
I sorted the problem. the long cooper pins were misaligned on the inside end of the connection block so i couldnt see anything wrong when looking from the side.
Rookie mistake..... though atleast the scrren wasnt buggered hahaha
Hi
After powering up a brand-new 3051(flange mounted wetleg) I noticed the screen does not show any indication(its blank). I can communicate with the instrument with Hart 375 communicator.
2 days earlier I unscrewed the screen and pulled it out to see if I could change the rotation of its...